Renal arteries: Renal a stenosis eval in transplanted kidney
| Specific Anatomic Region | Renal arteries | |
| Application | Renal artery stenosis evaluation in a transplanted kidney | |
| Author | Fishman EK | |
| Reference Source | Personal Communication | |
| Scanner Used | Siemens Plus4 Volume Zoom | |
| KV / mAs / Time per Rotation (sec) | 120 / 125 / .5 | |
| Detector Collimation (mm) | 4 x 1 | |
| Slice Thickness (mm) | 1.25 | |
| Data Reconstruction Interval (mm) | 1 | |
| Table Speed (mm per rotation) / Pitch | 6 / 6 | |
| Oral Contrast | 750 cc of water | |
| Contrast Volume and Type | 100-120 cc of Omnipaque 350 | |
| Injection Rate | 3 ml / sec | |
| Scan Delay (sec) | 30 sec | |
| 3D Technique Used | Volume rendering supplemented by MIPS. | |
| Comment: Non contrast scans are first obtained to localize the renal aa. Anastomosis. Contrast volume can be reduced to 80 cc if the patient has borderline renal function. | ||