Multidetector CT:
Vascular:

Renal arteries: Renal artery stenosis evaluation
Specific Anatomic Region Renal arteries
Application Renal artery stenosis evaluation
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 120 / 130 / .5 or 120 / 165 / .5
Detector Collimation (mm) 1
Slice Thickness (mm) 1.25
Data Reconstruction Interval (mm) 1
Table Speed (mm per rotation) / Pitch 6 / 6
Oral Contrast 750 cc of water
IV Contrast Volume and Type 100-120 cc of Omnipaque 350
Injection Rate 3 cc / sec
Scan Delay (sec) 25 sec
3D Technique Used Volume rendering
Comment: -A potentially even better protocol that we will begin looking at is using .5 mm detector, and a slice thickness of .5 mm, and reconstructing at .5 mm. Scan parameters 120/130/.5.