Pulmonary artery: R/O pulmonary embolism
| Specific Anatomic Region | Pulmonary artery | |
| Application | R/O pulmonary embolism | |
| Author | Fishman EK | |
| Reference Source | Personal Communication | |
| Scanner Used | Siemens Plus4 Volume Zoom | |
| KV / mAs / Time per Rotation (sec) | 140 / 100 / .5 | OR 140 / 100 / .5 |
| Detector Collimation (mm) | 2.5 | 1 |
| Slice Thickness (mm) | 3 | 1 |
| Data Reconstruction Interval (mm) | 2 | 1 |
| Table Speed (mm per rotation) / Pitch | 15 / 6 | 6 / 6 |
| Oral Contrast | N/A | |
| IV Contrast Volume and Type | 100-120 cc of Omnipaque 350 | |
| Injection Rate | 3 cc / sec | |
| Scan Delay (sec) | 20-25 sec | |
| 3D Technique Used | MPR or volume rendering (as needed) | |
| Comment: -MPR or volume rendering can be done in select cases to differentiate a node from a vessel. | ||