Multidetector CT:
Vascular:

Abdominal aorta: Evaluate known or suspected AAA
Specific Anatomic Region Abdominal aorta
Application Evaluate known or suspected AAA
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 120 / 165 / .5
Detector Collimation (mm) 1 or 2.5
Slice Thickness (mm) 1.25 or 3
Data Reconstruction Interval (mm) 1 or 2
Table Speed (mm per rotation) / Pitch 12.5 / 5 or 6 / 6
Oral Contrast 750 cc of water
IV Contrast Volume and Type 120 cc of Omnipaque 350
Injection Rate 3 cc / sec
Scan Delay (sec) 25 sec
3D Technique Used Volume rendering
Comment: -This protocol allows a compromise between length of study and 3D detail. When using 3 mm collimation, the entire abdominal aorta (and iliacs aa.) are scanned in 16-20 seconds.