Multidetector CT:
Spleen:

R/O splenic infarct
  Specific Anatomic Region Spleen
Application R/O splenic infarct
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 120 / 165 / .5
Detector Collimation (mm) 2.5
Slice Thickness (mm) 5
Data Reconstruction Interval (mm) 3
Table Speed (mm per rotation) / Pitch 12.5 / 5
  Oral Contrast 750 cc of water
  IV Contrast Volume and Type 100-120 cc    
  Injection Rate 2-3 cc
Scan Delay (sec) 50-60 sec
3D Technique Used N/A
Comment: -Delayed phase images are ideal, so that the splenic flow artifacts are not a problem.