Multidetector CT:
Pancreas:

Stage pancreatic mass (dual phase CT angio)
Specific Anatomic Region Pancreas
Application Stage pancreatic mass (dual phase CT angio)
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 120 / 125 / .5 or 120 / 150 / .5
Detector Collimation (mm) 1
Slice Thickness (mm) 1.25
Data Reconstruction Interval (mm) 1
Table Speed (mm per rotation) / Pitch 6 / 6
Oral Contrast 750 cc of water
IV Contrast Volume and Type 120 cc of Omnipaque 350
Injection Rate 3 cc / sec
Scan Delay (sec) 25 sec / 50 sec
3D Technique Used Volume rendering
Comment: -For larger patients, one can adjust the mAs upward, but be careful not to change it enough to have it change the pitch.