Multidetector CT:
Neuro:

Circle of Willis: Detection and assessment of intracranial aneurysm
Specific Anatomic Region Circle of Willis
Application Detection and assessment of intracranial aneurysm
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 120 / 109 / .5 or 120 / 200 / .5
Detector Collimation (mm) 1
Slice Thickness (mm) 1.25
Data Reconstruction Interval (mm) 1
Table Speed (mm per rotation) / Pitch 5 / 5 or 6 / 6
Oral Contrast N/A
IV Contrast Volume and Type 100-120 cc of Omnipaque 350
Injection Rate 3 cc / sec
Scan Delay (sec) 20 sec
3D Technique Used Volume rendering
Comment: -The key is to acquire all of the data in the arterial phase.