Circle of Willis: R/O aneurysm/evaluate aneurysm
| Specific Anatomic Region | Circle of Willis | |
| Application | R/O aneurysm/evaluate aneurysm | |
| Author | Fishman EK | |
| Reference Source | Personal Communication | |
| Scanner Used | Siemens Plus4 Volume Zoom | |
| KV / mAs / Time per Rotation (sec) | 120 / 109 / .5 or 120 / 165 / .5 | |
| Detector Collimation (mm) | .5 or 1 | |
| Slice Thickness (mm) | 1 or 1.25 | |
| Data Reconstruction Interval (mm) | 1 | |
| Table Speed (mm per rotation) / Pitch | 1.3 / 2.6 or 6 / 6 | |
| Oral Contrast | N/A | |
| IV Contrast Volume and Type | 100-120 cc of Omnipaque 350 | |
| Injection Rate | 3 cc / sec | |
| Scan Delay (sec) | 15-20 sec | |
| 3D Technique Used | Volume rendering | |
| Comment: -A key to this study is the quality of the dataset for 3D imaging. | ||