Diaphragm: Rupture/hernia
| Specific Anatomic Region | Diaphragm | |
| Application | Rupture/hernia | |
| Author | Fishman EK | |
| Reference Source | Personal Communication- (Lisa Reid) | |
| Scanner Used | Siemens Plus4 Volume Zoom | |
| KV / mAs / Time per Rotation (sec) | 140 / 120 / .5 | |
| Detector Collimation (mm) | 2.5 | |
| Slice Thickness (mm) | 3 | |
| Data Reconstruction Interval (mm) | 2 | |
| Table Speed (mm per rotation) / Pitch | 15.0 / 6 | |
| Oral Contrast | 500-750 cc of water 15 minutes prior to the study | |
| IV Contrast Volume and Type | 120 cc of Omnipaque 350 | |
| Injection Rate | 2-3 cc / sec | |
| Scan Delay (sec) | 30 sec | |
| 3D Technique Used | Volume rendering | |
| Comment: -In cases of suspected rupture, multiplanar images are ideal, especially in the sagittal plane. | ||