Suspected pulmonary embolism (classic)
| Specific Anatomic Region | Chest | |
| Application | Suspected pulmonary embolism (classic) | |
| Author | Fishman EK | |
| Reference Source | Personal Communication | |
| Scanner Used | Siemens Plus4 Volume Zoom | |
| KV / mAs / Time per Rotation (sec) | 140 / 100 / .5 | |
| Detector Collimation (mm) | 2.5 | |
| Slice Thickness (mm) | 3 | |
| Data Reconstruction Interval (mm) | 2 | |
| Table Speed (mm per rotation) / Pitch | 15 / 6 | |
| Oral Contrast | N/A | |
| IV Contrast Volume and Type | 120 cc of Omnipaque 350 | |
| Injection Rate | 3 cc / sec | |
| Scan Delay (sec) | 30 sec | |
| 3D Technique Used | MPR or volume rendering in select cases | |
| Comment: -This matches the classic technique used with single detector CT. A combination of 1 mm slice thickness of collimation may be considered for PE & is under investigation. | ||