Heart: Calcium scoring
| Specific Anatomic Region | Heart | |
| Application | Calcium scoring | |
| Author | Fishman EK | |
| Reference Source | Personal Communication | |
| Scanner Used | Siemens Plus4 Volume Zoom | |
| KV / mAs / Time per Rotation (sec) | 120 / 149 / .36 sec (partial rotation) | |
| Detector Collimation (mm) | 4x 2.5 | |
| Slice Thickness (mm) | 3 | |
| Data Reconstruction Interval (mm) | 3 | |
| Table Speed (mm per rotation) / Pitch | N/A- Sequential study | |
| Oral Contrast | N/A | |
| IV Contrast Volume and Type | N/A | |
| Injection Rate | N/A | |
| Scan Delay (sec) | N/A | |
| 3D Technique Used | N/A | |
| Comment: -The key to this study is obtained data during diastole. This is a gated study which is key to its accuracy. The actual cardiac calcification scoring is a post processing function (See other section on cardiac CT). | ||