Lung mass: R/O AVM; evaluate AVM
| Specific Anatomic Region | Lung mass | |
| Application | R/O AVM; evaluate AVM | |
| Author | Fishman EK | |
| Reference Source | Personal Communication | |
| Scanner Used | Siemens Plus4 Volume Zoom | |
| KV / mAs / Time per Rotation (sec) | 140 / 100 / .5 | |
| Detector Collimation (mm) | 1 | |
| Slice Thickness (mm) | 1.25 | |
| Data Reconstruction Interval (mm) | 1 | |
| Table Speed (mm per rotation) / Pitch | 6 / 6 | |
| Oral Contrast | N/A | |
| IV Contrast Volume and Type | 120 cc of Omnipaque 350 | |
| Injection Rate | 3 cc / sec | |
| Scan Delay (sec) | 30 sec | |
| 3D Technique Used | Volume rendering | |
| Comment: -3D volume rendering is ideal to detect the presence of an AVM as well as define its feeding and draining vessels. Axial images alone are not enough in these cases. | ||