Specific Anatomic Region |
Pelvis |
|
Application |
Cervical cancer |
|
Author |
Fishman EK |
|
Reference Source |
Personal Communication |
|
Scanner Used |
Siemens Plus4 Volume Zoom |
|
KV / mAs / Time per Rotation (sec) |
120 / 165 / .5 |
|
Detector Collimation (mm) |
2.5 |
|
Slice Thickness (mm) |
5 |
|
Data Reconstruction Interval (mm) |
5 |
|
Table Speed (mm per rotation) / Pitch |
12.5 / .5 |
|
Oral Contrast |
750-1000 cc of 3% Hypaque up to 2 hours prior to the study
|
|
IV Contrast Volume and Type |
120 cc of Omnipaque 350
|
|
Injection Rate |
2 cc / sec |
|
Scan Delay (sec) |
50-60 sec |
|
3D Technique Used |
N/A |
|
Comments:
- An optional protocol is to scan the abdomen as listed above, but
to modify the pelvis protocol to 140 / 210 / .5.
- We are also now doing dual phase studies (30 and 60 sec) to see
if this will improve staging of disease.
|