Specific Anatomic Region |
Shoulder |
|
Application |
Trauma to shoulder joint (includes scapula and humerus) |
|
Author |
Fishman EK |
|
Reference Source |
Personal Communication |
|
Scanner Used |
Siemens Plus4 Volume Zoom |
|
KV / mAs / Time per Rotation (sec) |
140 / 300 / 1.5 or 120 / 200 / .5 |
|
Detector Collimation (mm) |
1 |
|
Slice Thickness (mm) |
1.25 or 3 |
|
Data Reconstruction Interval (mm) |
1 |
|
Table Speed (mm per rotation) / Pitch |
4 / 4 or 6 / 6 |
|
Oral Contrast |
N/A |
|
IV Contrast Volume and Type |
N/A
|
|
Injection Rate |
N/A |
|
Scan Delay (sec) |
N/A |
|
3D Technique Used |
Volume rendering |
|
Comment: -This is a very impressive protocol (1.25 mm slice thickness). I have found that 3 mm thick sections reconstructed by 1 or 2 mm is also satisfactory. |