Specific Anatomic Region |
Bone/muscle |
|
Application |
Possible muscle abscess |
|
Author |
Fishman EK |
|
Reference Source |
Personal Communication |
|
Scanner Used |
Siemens Plus4 Volume Zoom |
|
KV / mAs / Time per Rotation (sec) |
120 / 165 / .5 |
|
Detector Collimation (mm) |
2.5 |
|
Slice Thickness (mm) |
3 or 5 |
|
Data Reconstruction Interval (mm) |
3 or 5 |
|
Table Speed (mm per rotation) / Pitch |
12.5 / 5 |
|
Oral Contrast |
N/A |
|
IV Contrast Volume and Type |
120 cc of Omnipaque 350
|
|
Injection Rate |
2-3 cc / sec |
|
Scan Delay (sec) |
40 sec |
|
3D Technique Used |
N/A |
|
Comment: -The key parameter is the differential enhancement of normal muscle & abscess. Without IV contrast, many abscesses would be missed. |