MDCT PROTOCOLS

Liver

Liver transplant evaluation (donor)Printer-ready protocol

Specific Anatomic Region Liver
Application Liver transplant evaluation (donor)
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 120 / 150 / .5
Detector Collimation (mm) 1
Slice Thickness (mm) 1.25
Data Reconstruction Interval (mm) 1 mm and 1 and 5 mm (venous phase only)
Table Speed (mm per rotation) / Pitch 6 / 6
Oral Contrast 750 cc of water
IV Contrast Volume and Type 120-150 cc of Omnipaque 350
Injection Rate 3 cc / sec
Scan Delay (sec) 25 / 50-60 sec
3D Technique Used Volume rendering
Comment: -This protocol is for the patient who is a liver donor. We obtain a set of venous phase images at 5 mm intervals to do volumetrics. Typically in adult to adult transplants, we do right lobe volume, and in adult to child transplants, we do lateral left lobe volumes.