MDCT PROTOCOLS

Kidney

Suspected renal mass/Staging renal mass Printer-ready protocol

Specific Anatomic Region Kidney
Application Suspected renal mass/Staging renal mass
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 120 / 165 / .5 (dual phase)
Detector Collimation (mm) 2.5 or 1
Slice Thickness (mm) 3 or 1.25
Data Reconstruction Interval (mm) 3-5 or 1
Table Speed (mm per rotation) / Pitch 12.5 / 5 or 6 / 6
Oral Contrast 750 cc of water
IV Contrast Volume and Type 120 cc of Omnipaque 350
Injection Rate 3 cc/sec
Scan Delay (sec) 30-40 sec/80 sec
3D Technique Used Volume rendering

Comments:

  • Two phases are routinely obtained in these patients. We obtain a corticomedullary phase and then a delayed phase (180 sec). Both phases use the same scanning parameters.
  • 1 mm collimation is used in order to create 3D reconstructions with optimal detail.