MDCT PROTOCOLS

Kidney

Hematuria Printer-ready protocol

Specific Anatomic Region Kidney
Application Hematuria
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 120 / 165 / .5
Detector Collimation (mm) 2.5
Slice Thickness (mm) 3-5
Data Reconstruction Interval (mm) 3-5
Table Speed (mm per rotation) / Pitch 12.5 / 5
Oral Contrast 750 cc of water
IV Contrast Volume and Type 125 ml of Omnipaque 350
Injection Rate 2-3 cc
Scan Delay (sec) 25 sec
3D Technique Used Volume rendering
Comment: -Non-contrast CT is followed by a late arterial phase imaging and a delayed phase for ideal 3D imaging. The use of early phase imaging is controversial in selected cases.