MDCT PROTOCOLS

GI

Stomach: Gastric mass (Virtual endoscopy) Printer-ready protocol

Specific Anatomic Region Stomach
Application Gastric mass (Virtual endoscopy)
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 120 / 125 / .5 or 120 / 165 / .5
Detector Collimation (mm) 1 or 2.5
Slice Thickness (mm) 1.25 or 3
Data Reconstruction Interval (mm) 1
Table Speed (mm per rotation) / Pitch 6 / 6 or12.5 / 5
Oral Contrast 750-1000 cc of water at the time of the study
IV Contrast Volume and Type 120 cc of Omnipaque 350
Injection Rate 3 cc/sec
Scan Delay (sec) 25-30 sec
3D Technique Used Volume rendering

Comments:

  • A key to this study is adequate gastric distension. We use water as the contrast agent of choice.
  • Rapid injection of IV contrast is critical for enhancing wall definition.