MDCT PROTOCOLS

Chest

Lung mass: R/O AVM; evaluate AVM Printer-ready protocol

Specific Anatomic Region Lung mass
Application R/O AVM; evaluate AVM
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 140 / 100 / .5
Detector Collimation (mm) 1
Slice Thickness (mm) 1.25
Data Reconstruction Interval (mm) 1
Table Speed (mm per rotation) / Pitch 6 / 6
Oral Contrast N/A
IV Contrast Volume and Type 120 cc of Omnipaque 350
Injection Rate 3 cc / sec
Scan Delay (sec) 30 sec
3D Technique Used Volume rendering
Comment: -3D volume rendering is ideal to detect the presence of an AVM as well as define its feeding and draining vessels. Axial images alone are not enough in these cases.