| Specific Anatomic Region |
Chest |
| Application |
Suspected pulmonary embolism (classic)
|
| Author |
Fishman EK |
| Reference Source |
Personal Communication |
| Scanner Used |
Siemens Plus4 Volume Zoom |
| KV / mAs / Time per Rotation
(sec) |
140 / 100 / .5 |
| Detector Collimation (mm) |
2.5 |
| Slice Thickness (mm) |
3 |
| Data Reconstruction Interval
(mm) |
2 |
| Table Speed (mm per rotation)
/ Pitch |
15 / 6 |
| Oral Contrast |
N/A |
| IV Contrast Volume and Type |
120 cc of Omnipaque 350 |
| Injection Rate |
3 cc / sec |
| Scan Delay (sec) |
30 sec |
| 3D Technique Used |
MPR or volume rendering in select
cases |
| Comment: -This matches the
classic technique used with single detector CT. A combination of 1 mm
slice thickness of collimation may be considered for PE & is under
investigation. |