MDCT PROTOCOLS

Chest

Dedicated HRCT (High resolution CT) Printer-ready protocol

Specific Anatomic Region Chest
Application Dedicated HRCT (High resolution CT)
Author Fishman EK
Reference Source Personal Communication
Scanner Used Siemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec) 140 / 120 / .75
Detector Collimation (mm) 1
Slice Thickness (mm) 1 or 1.25
Data Reconstruction Interval (mm) 5 or 10
Table Speed (mm per rotation) / Pitch 6 / 6
Oral Contrast N/A
IV Contrast Volume and Type N/A
Injection Rate N/A
Scan Delay (sec) N/A
3D Technique Used N/A

Comments:

  • This uses the B70 kernel (high resolution). By using .75 sec, we do oversampling of the data.
  • A protocol of 1mm collimations, 1.25 mm slice thickness, and 1mm reconstruction interval creates ideal 3D maps.