Sensation 64 Scanner: How it worksThomas Flohr, Siemens Medical Solutions The SOMATOM Sensation 64 makes use of a periodic motion of the focal spot in the longitudinal direction (z-flying focal spot) to double the number of simultaneously acquired slices with the goal of improved longitudinal resolution and elimination of spiral artifacts. By permanent electromagnetic deflection of the electron beam in the X-ray tube the focal spot is wobbled between two different positions on the anode plate. Due to the anode angle of 7° this translates into a motion both in the radial direction and in the z-direction. The radial motion is a side-effect which is taken care of by the image reconstruction algorithms. The amplitude of the periodic z-motion is adjusted such that two subsequent readings are shifted by half a collimated slice width in the patient's longitudinal direction, see below.
Therefore,
the measurement rays of two subsequent readings with collimated slice-width
0.6 mm interleave in the z-direction, and every two 32-slice readings
are combined to one 64-slice projection with a sampling distance of
0.3 mm. With this technique, 64 overlapping 0.6 mm slices per rotation
are acquired. The z-coverage is 32x0.6 mm = 19.2 mm, and the sampling
scheme is identical to that of a 64x0.3 mm detector. This fine sampling
is the reason for the improved spatial resolution and the elimination
of spiral artifacts. The improved sampling is obtained at any spiral
pitch. Hence, resolution is improved and spiral artifacts are eliminated
at any pitch. The improved sampling is furthermore not restricted to
the iso-center, but is maintained in a wide range of the scan field
of view (SFOV), see the figure above. This is a major difference to
conventional approaches which attempt to improve longitudinal sampling
by the choice of optimized small pitch values (so-called "High
Quality" pitches), so that data acquired in different rotations
interleave in the z-direction. In this case, a sampling distance of
half the collimated slice width can be achieved close to iso-center
only, see below, and improved resolution will only be achieved close
to iso-center, which is a clinical drawback. Free selection of the pitch
is not possible, and the user is restricted to special modes with low
table feed and low volume coverage. To sum up: The z-flying focal spot technique improves longitudinal resolution and eliminates windmill artifacts at any pitch. Even the most demanding clinical applications can be performed at maximum pitch without degradation of image quality or resolution. This is a major difference to conventional CT-systems which claim to have a larger detector coverage but are restricted in table feed ("High Quality" Modes), if good image quality is required. |