VascularRenal arteries: Renal artery stenosis evaluation |
|
Specific Anatomic Region |
Renal Arteries |
|
|
Application |
Renal artery stenosis evaluation |
|
|
Author |
Beregi JP et al. |
|
|
Reference Source |
AJR 1996; 167:495-501 |
|
|
Scanner Used |
Siemens Somatom Plus-S |
|
|
Injection Rate |
4ml/sec |
|
|
Contrast Volume and Type |
120ml of nonionic 300mg I/ml |
|
|
Area Scanned |
Area of renal aa first defined on noncontrast scans. This area only is scanned. |
|
|
Scan Delay |
Determined by test dose |
|
|
Length of Spiral (time) |
20sec |
|
|
Slice Thickness |
3mm |
|
|
Table Speed/Pitch |
3mm/sec |
|
|
Reconstruction Interval |
1mm |
|
|
3D Technique Used |
SSD and MIP |
|
|
Comment: The authors reported excellent results but for the "main" renal aa as this protocol covers but 6cm. This protocol took about 25 minutes per patient according to the authors. |
||