VascularRenal arteries: Renal artery stenosis evaluation |
|
Specific Anatomic Region |
Renal arteries |
|
|
Application |
Renal artery stenosis evaluation |
|
|
Author |
Galanski M et al. |
|
|
Reference Source |
Radiology 1993; 189:185-192 |
|
|
Scanner Used |
Siemens Somatom Plus or Plus-S |
|
|
Injection Rate |
3-5ml/sec (choice 3.5ml/sec) |
|
|
Contrast Volume and Type |
100-150ml of Iopromide-300 |
|
|
Area Scanned |
Entire length of the kidneys |
|
|
Scan Delay |
Test-dose timing (usually 20-30sec) |
|
|
Length of Spiral (time) |
32sec |
|
|
Slice Thickness |
2mm |
|
|
Table Speed/Pitch |
2mm/sec with a pitch of 1 |
|
|
Reconstruction Interval |
1-2mm |
|
|
3D Technique Used |
Gradient shaded-surface rendering (with threshold of 130-220 HU) and MIP |
|
|
Comment: Test dosing to optimize timing was used in this study. This is similar to work by Rubin et al. but they inject at 5cc/sec. A key in these cases is to look for accessory renal arteries which increase the length of coverage needed. |
||