VascularCarotid arteries: Carotid artery stenosis |
|
Specific Anatomic Region |
Carotid Arteries |
|
|
Application |
Carotid artery stenosis |
|
|
Author |
Kuszyk B.S., Beauchamp, N.J. and Fishman, E.K. (1998). |
|
|
Reference Source |
Neurovascular applications of CT angiography. Semin US CT MR, 19(5), 394-404. |
|
|
Scanner Used |
Siemens Somatom Plus 4 |
|
|
Injection Rate |
3cc/sec |
|
|
Contrast Volume and Type |
120cc of Omnipaque 350 |
|
|
Area Scanned |
From C4 to base of skull |
|
|
Scan Delay |
18sec |
|
|
Length of Spiral (time) |
25-30sec |
|
|
Slice Thickness |
2mm |
|
|
Table Speed/Pitch |
4mm/sec or a pitch of 2 |
|
|
Reconstruction Interval |
1mm |
|
|
3D Technique Used |
Volume rendering with MIP as a supplement |
|
|
Comment: The key is to edit out obscuring structures in order to define the carotid aa without overlapping structures which may interfere with accuracy. |
||