Chest Printer-ready protocol

Suspected pulmonary embolism

Specific Anatomic Region

Chest

Application

Suspected pulmonary embolism

Author

Fishman EK

Reference Source

Personal communication

Scanner Used

Siemens Somatom Plus 4

Injection Rate

3 ml/sec

Contrast Volume and Type

120ml of Omnipaque 350

Area Scanned

Level above aortic arch through lung bases

Scan Delay

25 sec

Length of Spiral (time)

30-35 sec

Slice Thickness

3 mm

Table Speed/Pitch

5 mm/sec

Reconstruction Interval

2 mm

3D Technique Used

Volume Rendering or MPR

Comment: We will scan from the lung bases upward to take advantage of better patient cooperation with breathholding in this critical region. Recent work suggests that Visipaque may be valuable due to better opacification of smaller vessels.